Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Published July 3, 2024 · Recall initiated: May 13, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
SIGHT DIAGNOSTICS LTD
Distributor
Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.
Lot
REF: TK1, UDI-DI: B787TK10, Lot Code: 10114
Risk classification
Class II
Area (FDA)
device
Country
Israel