Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Published July 3, 2024 · Recall initiated: May 13, 2024

Summary

FDA records medical device Class II recall Z-2173-2024, published July 3, 2024 and initiated May 13, 2024. Recalling firm: SIGHT DIAGNOSTICS LTD; Status: Ongoing; Reported quantity: 7,450 units (2,800 US, 4,650 OUS); Lot: REF: TK1, UDI-DI: B787TK10, Lot Code: 10114. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
SIGHT DIAGNOSTICS LTD
Distributor
Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.
Lot
REF: TK1, UDI-DI: B787TK10, Lot Code: 10114
Risk classification
Class II
Area (FDA)
device
Country
Israel

Related recalls

Questions about this recall

Was Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer. recalled?

Yes. FDA published recall Z-2173-2024 on July 3, 2024. Reported hazard: One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

What should I do if I have this product?

Voluntary: Firm initiated · Letter