Back to search
FDAFDA · Food and drugs·Recall number Z-2175-2024

The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Published July 3, 2024 · Recall initiated: May 17, 2024

Summary

FDA records medical device Class II recall Z-2175-2024, published July 3, 2024 and initiated May 17, 2024. Recalling firm: Maquet Cardiovascular, LLC; Status: Ongoing; Reported quantity: 430,037 units; Lot: 1. Model Number: VH-4000. UDI-DI: 00607567700406. All lot numbers within expiration. 2. Model Number: VH-4001. UDI-DI: 00607567700451. All lot numbers within expiration. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
Lot
1. Model Number: VH-4000. UDI-DI: 00607567700406. All lot numbers within expiration. 2. Model Number: VH-4001. UDI-DI: 00607567700451. All lot numbers within expiration.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization. recalled?

Yes. FDA published recall Z-2175-2024 on July 3, 2024. Reported hazard: Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

What should I do if I have this product?

Voluntary: Firm initiated · Letter