System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Published August 15, 2012 · Recall initiated: July 26, 2012
Summary
FDA records medical device Class II recall Z-2177-2012, published August 15, 2012 and initiated July 26, 2012. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 992 units; Lot: catalog number: 801763 and serial numbers: 0011-0335, 0337-0610,1001-1034, 1037-1049,1100,1103-1106,1109-1450, and 1452. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Japan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
- Lot
- catalog number: 801763 and serial numbers: 0011-0335, 0337-0610,1001-1034, 1037-1049,1100,1103-1106,1109-1450, and 1452.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPMJuly 2, 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974April 10, 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, SterileDecember 23, 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19REAugust 7, 2019
- FDAStryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument,August 15, 2012
- FDASystem 1 base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporealAugust 15, 2012
Questions about this recall
Was System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment. recalled?
Yes. FDA published recall Z-2177-2012 on August 15, 2012. Reported hazard: Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re
What should I do if I have this product?
Voluntary: Firm initiated · Letter