System 1 base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Published August 15, 2012 · Recall initiated: July 26, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Japan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
Lot
catalog number: 801764 and serial numbers: 0006-0066, 0100-0311, 0313-0322, 1001-1027, 1100-1434, 1436, 1440, and 1441.