Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

Published August 6, 2025 · Recall initiated: June 23, 2025

Summary

FDA records medical device Class II recall Z-2178-2025, published August 6, 2025 and initiated June 23, 2025. Recalling firm: Steris Corporation; Status: Ongoing; Reported quantity: 59 units; Lot: UDI: 00724995171155: Serial numbers: 0416920048, 0417820084, 0417820088, 0417820096, 0417820110, 0417820092, 0417820100, 0418420047, 0419620004, 0419520013, 0428620109, 0428620112, 0430020191, 0431720044, 0435120043, 0435120047, 0435120051, 0401321042, 0413221047, 0413221199, 0413721008, 0413721012, 0417221023, 0417221027, 0417221031, 0420021018, 0420021022, 0420021026, 0420721021, 0429720003, 0420921073, 0426321186, 0426321190, 0427121121, 0427121125, 0427321012, 0427321002, 0427321005, 0413921017, 0434121095, 0434121098, 0434121101, 0434121104, 0434121107, 0434121110, 0434121113, 0400722081, 0406022021, 0406022025, 0406022034, 0406022038, 0406022017, 0406022014, 0406022011, 0417122078, 0417122081, 0417122084, 0417122087, 0417122090. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steris Corporation
Distributor
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Lot
UDI: 00724995171155: Serial numbers: 0416920048, 0417820084, 0417820088, 0417820096, 0417820110, 0417820092, 0417820100, 0418420047, 0419620004, 0419520013, 0428620109, 0428620112, 0430020191, 0431720044, 0435120043, 0435120047, 0435120051, 0401321042, 0413221047, 0413221199, 0413721008, 0413721012, 0417221023, 0417221027, 0417221031, 0420021018, 0420021022, 0420021026, 0420721021, 0429720003, 0420921073, 0426321186, 0426321190, 0427121121, 0427121125, 0427321012, 0427321002, 0427321005, 0413921017, 0434121095, 0434121098, 0434121101, 0434121104, 0434121107, 0434121110, 0434121113, 0400722081, 0406022021, 0406022025, 0406022034, 0406022038, 0406022017, 0406022014, 0406022011, 0417122078, 0417122081, 0417122084, 0417122087, 0417122090.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems recalled?

Yes. FDA published recall Z-2178-2025 on August 6, 2025. Reported hazard: The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter