FDAFDA · Food and drugs·Recall number Z-2210-2014
Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.
Published August 20, 2014 · Recall initiated: July 10, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medical Positioning Inc
- Distributor
- Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
- Lot
- Serial number--802722, 802368, 803147, 804007, 804132, 803195, 802286, 802303, 802782, 803112, 803200, 803274, 803455, 803468, 803883, 803961 and 803960.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States