Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used to conduct syncope examinations.

Published August 20, 2014 · Recall initiated: July 10, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medical Positioning Inc
Distributor
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
Lot
Serial number--802971, 803106, 803356, 803588, 803616, 803854, 802781, 802783, 802780, 802784, 802421, 802861, 802939, 802949, 802924, 803867, 803926, 802809 and 804118.
Risk classification
Class II
Area (FDA)
device
Country
United States