Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 1 (part # 700-020-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***single-port manifold for the Stryker Neptune 2 (part # 702-025-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 2 (part # 702-020-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump.

Published September 25, 2013 · Recall initiated: August 16, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol for reprocessing the used manifolds.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Zeppessis Reprocessing, LLC
Distributor
Distributed in the states of California, Idaho, Oregon, and Washington.
Lot
-Single-port manifold for the Stryker Neptune 1 (part # 700-015-000); -Four-port manifold for the Stryker Neptune 1 (part # 700-020-000); -Single-port manifold for the Stryker Neptune 2 (part # 702-025-000); -Four-port manifold for the Stryker Neptune 2 (part # 702-020-000).
Risk classification
Class II