FDAFDA · Food and drugs·Recall number Z-2213-2013

Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers

Published September 25, 2013 · Recall initiated: May 10, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path. Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Rauland-Borg Corp
Distributor
Worldwide distribution - US: (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NE, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and WV., and the countries of : Australia, Canada, Columbia, Lebanon, Mexico and Saudi Arabia.
Lot
1) Product Code 351000 (Branch Regional Controller); 2) Product Code 352000 (Corridor Light); 3) Product Code 352020 (Domeless Controller)
Risk classification
Class II
Area (FDA)
device
Country
United States