Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers
Published September 25, 2013 · Recall initiated: May 10, 2013
Summary
FDA records medical device Class II recall Z-2213-2013, published September 25, 2013 and initiated May 10, 2013. Recalling firm: Rauland-Borg Corp; Status: Terminated; Reported quantity: Product Code 351000: 385 units; Product Codes 352000 and 352020: 14,808 units (total); Lot: 1) Product Code 351000 (Branch Regional Controller); 2) Product Code 352000 (Corridor Light); 3) Product Code 352020 (Domeless Controller). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path. Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Rauland-Borg Corp
- Distributor
- Worldwide distribution - US: (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NE, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and WV., and the countries of : Australia, Canada, Columbia, Lebanon, Mexico and Saudi Arabia.
- Lot
- 1) Product Code 351000 (Branch Regional Controller); 2) Product Code 352000 (Corridor Light); 3) Product Code 352020 (Domeless Controller)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAZeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste ManagementSeptember 25, 2013
- FDACypher MIS Screw System, Biomet Spine Parsippany, NJ 07054. The Cypher MIS Screw System is a non-cervicalSeptember 25, 2013
- FDADevice Name: Terino Square Chin-Style II-Medium; Catalog Number: TSCII-M. To augment or reconstructSeptember 25, 2013
- FDABIORAPTOR 2.9 mm Suture Anchor with 2 ULTRABRAID Suture; Part Number: 72200775 Biodegradable suture anchorSeptember 25, 2013
- FDABrilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/NSeptember 25, 2013
- FDABIORAPTOR 2.9 mm Suture Anchor with 1 ULTRABRAID Suture; Part Number: 72200774 Biodegradable suture anchorSeptember 25, 2013
Questions about this recall
Was Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers recalled?
Yes. FDA published recall Z-2213-2013 on September 25, 2013. Reported hazard: A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path. Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail