cobas¿ Liat¿ Analyzer

Published July 27, 2016 · Recall initiated: May 9, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Roche Molecular Systems, Inc.
Distributor
Distributed in NC and WA.
Lot
US-IVD: M1-E-00383 & M1-E-00434
Risk classification
Class II
Area (FDA)
device
Country
United States