Back to search
FDAFDA · Food and drugs·Recall number Z-2217-2020

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X50MM, Item Number 810M5550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Published June 10, 2020 · Recall initiated: May 5, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer Biomet, Inc.
Distributor
US nationwide distribution.
Lot
UDI: (01)00887868324451; Lot Numbers: C19H0010 C19I0008 C19J0012 C19K0008 C20A0005 SBM108406 SBM118988 SBM122122
Risk classification
Class II
Area (FDA)
device
Country
United States