Back to search
FDAFDA · Food and drugs·Recall number Z-2222-2020

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X55MM, Item Number 810M6555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Published June 10, 2020 · Recall initiated: May 5, 2020

Summary

FDA records medical device Class II recall Z-2222-2020, published June 10, 2020 and initiated May 5, 2020. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 3235 total; Lot: UDI: (01)00887868324888; Lot Numbers: C19I0001 C19I0010 C19J0020 C19L0003 C20A0018 SBM108412 SBM118994. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer Biomet, Inc.
Distributor
US nationwide distribution.
Lot
UDI: (01)00887868324888; Lot Numbers: C19I0001 C19I0010 C19J0020 C19L0003 C20A0018 SBM108412 SBM118994
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X55MM, Item Number 810M6555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients. recalled?

Yes. FDA published recall Z-2222-2020 on June 10, 2020. Reported hazard: There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

What should I do if I have this product?

Voluntary: Firm initiated · Letter