Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.
Published August 29, 2012 · Recall initiated: July 27, 2012
Summary
FDA records medical device Class II recall Z-2224-2012, published August 29, 2012 and initiated July 27, 2012. Recalling firm: Biomet U.K., Ltd.; Status: Terminated; Reported quantity: 119 units; Lot: Catalog Number: 32-422801 and Lot Number Identification: ZB111101, ZBll1201, ZB120101, ZB120501, ZB111102, and ZB120201. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet U.K., Ltd.
- Distributor
- Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.
- Lot
- Catalog Number: 32-422801 and Lot Number Identification: ZB111101, ZBll1201, ZB120101, ZB120501, ZB111102, and ZB120201.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAOxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size F Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Phase 3 Tibial Template Right Medial Size B Product Usage: Instruments for use withMarch 1, 2017
- FDAThe Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produceAugust 29, 2012
- FDAUniCel DxH Slidemaker Stainer Instrument System, PN 775222. The DxH Slidemaker Stainer is a fully automatedAugust 29, 2012
Questions about this recall
Was Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component. recalled?
Yes. FDA published recall Z-2224-2012 on August 29, 2012. Reported hazard: A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension
What should I do if I have this product?
Voluntary: Firm initiated · Letter