FDAFDA · Food and drugs·Recall number Z-2224-2020
Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X40MM, Item Number 810M6540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.
Published June 10, 2020 · Recall initiated: May 5, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- US nationwide distribution.
- Lot
- UDI: (01)00887868324857; Lot Numbers: C19H0014 C19I0015 C19J0013 C19K0011 C19L0012 C19L0015 C19L0016 SBM108409 SBM118991 SBM122159
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.