Dimension Vista® Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.
Published July 27, 2016 · Recall initiated: May 17, 2016
Summary
FDA records medical device Class II recall Z-2241-2016, published July 27, 2016 and initiated May 17, 2016. Recalling firm: CSL Behring GmbH; Status: Terminated; Reported quantity: 898 units; Lot: Lot Number 15243MA Mfg Date: 08/31/2015 Exp Date: 05/31/2016 Lot Number 15243MB Mfg Datge: 08/31/2015 Exp Date: 05/31/2016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CSL Behring GmbH
- Distributor
- Distributed to: AL, AR, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WI, WV
- Lot
- Lot Number 15243MA Mfg Date: 08/31/2015 Exp Date: 05/31/2016 Lot Number 15243MB Mfg Datge: 08/31/2015 Exp Date: 05/31/2016
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDASIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/ProductDecember 6, 2017
- FDA2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the followingJuly 27, 2016
- FDAConvenience Kits from Centurion that contain a Medtronic/Covidien Light Glove ED SUTURE KIT VAGINAL DELIVERYJuly 27, 2016
- FDAT100-022WF, IV Set with 2 Y-Sites, 0.22 micron filter, Back Check Valve, & FP Clip. Packaged in a Tyvek pouch,July 27, 2016
- FDAAnspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones ofJuly 27, 2016
- FDAT105-000WC IV Set w/2 Y-Sites, BC Vlv & Filter Cap. Packaged in a Tyvek pouch, with 50 pouches per case.July 27, 2016
Questions about this recall
Was Dimension Vista® Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System. recalled?
Yes. FDA published recall Z-2241-2016 on July 27, 2016. Reported hazard: Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter