FDAFDA · Food and drugs·Recall number Z-2243-2026
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Published June 3, 2026 · Recall initiated: April 24, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- GE Medical Systems Information Technologies Inc
- Distributor
- Worldwide distribution.
- Lot
- UDI none, Serial number or Sales Order number SS921120014SA, SS921120017SA, SS921120021SA, SS921120018SA, SS921120020SA, SS920390001SA, SS921120007SA, SS921120008SA, SS921120012SA, SS921120016SA, SS920500001SA, SS920500002SA, SS921120009SA, SS921120010SA.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States