Back to search
FDAFDA · Food and drugs·Recall number Z-2246-2018

Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set

Published June 27, 2018 · Recall initiated: March 5, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Synthes (USA) Products LLC
Distributor
Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
Lot
FL00860
Risk classification
Class II
Area (FDA)
device
Country
United States