Back to search
FDAFDA · Food and drugs·Recall number Z-2246-2018

Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set

Published June 27, 2018 · Recall initiated: March 5, 2018

Summary

FDA records medical device Class II recall Z-2246-2018, published June 27, 2018 and initiated March 5, 2018. Recalling firm: Synthes (USA) Products LLC; Status: Terminated; Reported quantity: 4; Lot: FL00860. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Synthes (USA) Products LLC
Distributor
Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
Lot
FL00860
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set recalled?

Yes. FDA published recall Z-2246-2018 on June 27, 2018. Reported hazard: Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

What should I do if I have this product?

Voluntary: Firm initiated · Letter