Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body.
Published September 19, 2012 · Recall initiated: July 10, 2012
Summary
FDA records medical device Class II recall Z-2256-2012, published September 19, 2012 and initiated July 10, 2012. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 254 total in the US; Lot: Part Numbers: 5555000-3, 5555000-4, 5555000-5, 5555000-6. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution -- USA (nationwide) Distribution
- Lot
- Part Numbers: 5555000-3, 5555000-4, 5555000-5, 5555000-6.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAOptima XR220amx Mobile general purpose radiographic imaging of the human head and body.September 19, 2012
- FDABard Foley procedural trays packaged in breather bags. The product is packaged ten units in each shipper box,September 19, 2012
Questions about this recall
Was Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body. recalled?
Yes. FDA published recall Z-2256-2012 on September 19, 2012. Reported hazard: Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter