Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19; d) ES-12x13x15; e) ES-12x15x17; f) ES-12x16x19; g) ES-15x13x16
Published June 27, 2018 · Recall initiated: April 13, 2018
Summary
FDA records medical device Class II recall Z-2264-2018, published June 27, 2018 and initiated April 13, 2018. Recalling firm: TriMed Inc.; Status: Terminated; Reported quantity: 706; Lot: Lot #: a) 00A5, 26406, R889, 03UJ, 003UJ, 050514, 10505-14, 264-06, 004P, 26404, 10505-14, 226001, 264, 26401, 22600; b) 004S; c) 004R; d) 050518, 10505-18, 021G, 021F, 0505018, 11788, 080818, 178801, 17881, 316801; e) R895, 050519, 10505-19, 2012BJ, 004V, 009D, 0505019, 178901, 180502, 17890, 1178901, 00AB, 0090, 0091, 78901; f) 009E, 11264, R896, 004W, 183701, 11837, 050520, 10505-20, 178901, 10505-20, 083701, 118337, 08370, 26407, 264-07; g) 03UR, 26408, 264-08, 050521, 26406, 264-06, 1050521, 179001, 10505-21, 179002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- TriMed Inc.
- Distributor
- U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
- Lot
- Lot #: a) 00A5, 26406, R889, 03UJ, 003UJ, 050514, 10505-14, 264-06, 004P, 26404, 10505-14, 226001, 264, 26401, 22600; b) 004S; c) 004R; d) 050518, 10505-18, 021G, 021F, 0505018, 11788, 080818, 178801, 17881, 316801; e) R895, 050519, 10505-19, 2012BJ, 004V, 009D, 0505019, 178901, 180502, 17890, 1178901, 00AB, 0090, 0091, 78901; f) 009E, 11264, R896, 004W, 183701, 11837, 050520, 10505-20, 178901, 10505-20, 083701, 118337, 08370, 26407, 264-07; g) 03UR, 26408, 264-08, 050521, 26406, 264-06, 1050521, 179001, 10505-21, 179002
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM NailMay 22, 2024
- FDATriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM NailMay 22, 2024
- FDAVolar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W,February 7, 2024
- FDATriMed Hex Driver 1.2, 55mm AO Cannulated, REF HXDRIVR-1.2C AO, nonsterile, Rx. The firm name on the label isNovember 20, 2019
- FDAHoney Oat Bran Muffin, 4-pack, packaged in a clear clamshell container, net wt. 13.6 oz., UPC 8 40600 10954 9July 4, 2018
- FDANitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22June 27, 2018
Questions about this recall
Was Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19; d) ES-12x13x15; e) ES-12x15x17; f) ES-12x16x19; g) ES-15x13x16 recalled?
Yes. FDA published recall Z-2264-2018 on June 27, 2018. Reported hazard: During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
What should I do if I have this product?
Voluntary: Firm initiated · Letter