Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.
Published September 5, 2012 · Recall initiated: November 1, 2010
Summary
FDA records medical device Class II recall Z-2265-2012, published September 5, 2012 and initiated November 1, 2010. Recalling firm: Canon Inc.; Status: Terminated; Reported quantity: 468 units; Lot: Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Canon Inc.
- Distributor
- Nationwide Distribution (USA)
- Lot
- Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Related recalls
- FDACanon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to beJune 5, 2013
- FDAsyngo Imaging XS. Radiological image processing system.September 5, 2012
- FDAToshiba CT systems, Asteion and Aquilion, that are equipped with a PC based computer, or to those olderSeptember 5, 2012
- FDACisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured andSeptember 5, 2012
- FDACaviCide, Part Numbers: 13-1025, 13-1000, 13-1024, MC-1000. Alternate brands: EnviroCide, part number:September 5, 2012
- FDAON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMANDSeptember 5, 2012
Questions about this recall
Was Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. recalled?
Yes. FDA published recall Z-2265-2012 on September 5, 2012. Reported hazard: Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone