Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

Published September 5, 2012 · Recall initiated: November 1, 2010

Summary

FDA records medical device Class II recall Z-2265-2012, published September 5, 2012 and initiated November 1, 2010. Recalling firm: Canon Inc.; Status: Terminated; Reported quantity: 468 units; Lot: Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Canon Inc.
Distributor
Nationwide Distribution (USA)
Lot
Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA.
Risk classification
Class II
Area (FDA)
device
Country
Japan

Related recalls

Questions about this recall

Was Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. recalled?

Yes. FDA published recall Z-2265-2012 on September 5, 2012. Reported hazard: Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone