DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

Published August 13, 2025 · Recall initiated: June 25, 2025

Summary

FDA records medical device Class II recall Z-2267-2025, published August 13, 2025 and initiated June 25, 2025. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 1 system; Lot: UDI-DI: 15099590369231; Serial Numbers: 608109. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
Worldwide distribution - US Nationwide.
Lot
UDI-DI: 15099590369231; Serial Numbers: 608109
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457 recalled?

Yes. FDA published recall Z-2267-2025 on August 13, 2025. Reported hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter