DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Published August 13, 2025 · Recall initiated: June 25, 2025
Summary
FDA records medical device Class II recall Z-2267-2025, published August 13, 2025 and initiated June 25, 2025. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 1 system; Lot: UDI-DI: 15099590369231; Serial Numbers: 608109. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter, Inc.
- Distributor
- Worldwide distribution - US Nationwide.
- Lot
- UDI-DI: 15099590369231; Serial Numbers: 608109
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDADxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513October 21, 2026
- FDAIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345,October 21, 2026
- FDAUIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,August 13, 2025
- FDADxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456August 13, 2025
Questions about this recall
Was DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457 recalled?
Yes. FDA published recall Z-2267-2025 on August 13, 2025. Reported hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter