CS-7, Model No. A4C1
Published July 4, 2018 · Recall initiated: May 9, 2018
Summary
FDA records medical device Class II recall Z-2281-2018, published July 4, 2018 and initiated May 9, 2018. Recalling firm: Konica Minolta Medical Imaging USA, Inc.; Status: Terminated; Lot: All version numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Konica Minolta Medical Imaging USA, Inc.
- Distributor
- Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.
- Lot
- All version numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARandox Liquid Assayed Specific Protein Control - Level 2 (SP Control 2), Cat. no. PS2683July 4, 2018
- FDAABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction withJuly 4, 2018
- FDABrilliance Big Bore Radiology 728244July 4, 2018
- FDAArtis Q.zen floor, Material no. 10848353, for angiography and whole body radiographic/fluoroscopic procedures.July 4, 2018
- FDAWeis Signature Collection No Salt Added White Corn Tortilla Chips; Net Wt 11 oz; Manufactured For WeisJuly 4, 2018
- FDASonimage HS1 Ultrasound Kit AC adapter, Product Number: A8AR, UDI: 04560141941588April 4, 2018
Questions about this recall
Was CS-7, Model No. A4C1 recalled?
Yes. FDA published recall Z-2281-2018 on July 4, 2018. Reported hazard: Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.
What should I do if I have this product?
Voluntary: Firm initiated · Letter