Back to search
FDAFDA · Food and drugs·Recall number Z-2281-2018

CS-7, Model No. A4C1

Published July 4, 2018 · Recall initiated: May 9, 2018

Summary

FDA records medical device Class II recall Z-2281-2018, published July 4, 2018 and initiated May 9, 2018. Recalling firm: Konica Minolta Medical Imaging USA, Inc.; Status: Terminated; Lot: All version numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Konica Minolta Medical Imaging USA, Inc.
Distributor
Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.
Lot
All version numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CS-7, Model No. A4C1 recalled?

Yes. FDA published recall Z-2281-2018 on July 4, 2018. Reported hazard: Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

What should I do if I have this product?

Voluntary: Firm initiated · Letter