FDAFDA · Food and drugs·Recall number Z-2284-2012

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Published September 5, 2012 · Recall initiated: August 10, 2012

Summary

FDA records medical device Class II recall Z-2284-2012, published September 5, 2012 and initiated August 10, 2012. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 64 units; Lot: Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steris Corporation
Distributor
Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.
Lot
Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor. recalled?

Yes. FDA published recall Z-2284-2012 on September 5, 2012. Reported hazard: STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

What should I do if I have this product?

Voluntary: Firm initiated · Visit