Back to search
FDAFDA · Food and drugs·Recall number Z-2293-2016

MOSAIQ Oncology Information System

Published August 3, 2016 · Recall initiated: July 15, 2016

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It is possible that a change to an Order Set will not be saved in the current open Care Plan. This can result in a wrong chemo dose in the order.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Elekta, Inc.
Distributor
Worldwide distribution. US nationwide, Antigua and Barbuda, Australia, Austria, Bahamas, Bahrain, Canada, Czech Republic, Germany, Hungary, South Africa, Spain, Sweden, Turkey, and United Kingdom
Lot
Versions 2.50.05 and higher
Risk classification
Class II
Area (FDA)
device
Country
United States