BD Vacutainer PST Gel and Lithium HeparinN (LH) 56 Units Blood Collection Tubes, Catalog Number 367960
Published August 28, 2019 · Recall initiated: June 12, 2019
Summary
FDA records medical device Class II recall Z-2301-2019, published August 28, 2019 and initiated June 12, 2019. Recalling firm: Becton Dickinson & Company; Status: Terminated; Reported quantity: 1,436,228,004 total; Lot: UDI: (01)30382903679608. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- Worldwide Distribution, including US Nationwide
- Lot
- UDI: (01)30382903679608
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD Vacutainer Lithium HeparinN (LH) 95 USP Units Blood Collection Tubes, Catalog Number 367886August 28, 2019
- FDABD Vacutainer PST Gel and Lithium HeparinN (LH) 126 Units Blood Collection Tubes, Catalog Number 367964August 28, 2019
Questions about this recall
Was BD Vacutainer PST Gel and Lithium HeparinN (LH) 56 Units Blood Collection Tubes, Catalog Number 367960 recalled?
Yes. FDA published recall Z-2301-2019 on August 28, 2019. Reported hazard: False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
What should I do if I have this product?
Voluntary: Firm initiated · Letter