Back to search
FDAFDA · Food and drugs·Recall number Z-2305-2019

BD Vacutainer PST Gel and Lithium HeparinN (LH) 126 Units Blood Collection Tubes, Catalog Number 367964

Published August 28, 2019 · Recall initiated: June 12, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Becton Dickinson & Company
Distributor
Worldwide Distribution, including US Nationwide
Lot
UDI: (01)30382903679646
Risk classification
Class II
Area (FDA)
device
Country
United States