FDAFDA · Food and drugs·Recall number Z-2313-2024
VidiaPort carrying arm, mount for auxiliary products, model numbers: a) 4028110 (Ceiling Single Surgical Light); b) 4028210 (Ceiling Duo Surgical Lights); c) 4028310 (Ceiling Trio Surgical Lights)
Published July 17, 2024 · Recall initiated: June 4, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide distribution.
- Lot
- a) UDI/DI 00887761968523, Serial Numbers: 108664155, 108664156, 108669683, 108669684, 108673461, 108696617, 108726885, 108731089, 108746956, 108754496, 108754497, 108821898, 108835099, 108907603, 108909646, 108917888, 108919439, 108919440, 108919441, 108961413, 108961414, 108970453, 108970454; b) UDI/DI 00887761968462, Serial Numbers: 108645027, 108645028, 108645031, 108648403, 108652874, 108664909, 108664911, 108665618, 108669680, 108669681, 108669682, 108671489, 108671494, 108671495, 108671496, 108671497, 108671498, 108671499, 108671500, 108674055, 108674056, 108674057, 108674058, 108676619, 108676620, 108676621, 108677352, 108677353, 108678974, 108678975, 108680874, 108680875, 108680877, 108683306, 108692868, 108696905, 108698871, 108712900, 108713005, 108724511, 108724512, 108726889, 108726890, 108726969, 108726970, 108729819, 108729820, 108729821, 108729822, 108740036, 108740037, 108741241, 108741242, 108741243, 108741244, 108741245, 108741246, 108746957, 108746958, 108750621, 108750622, 108750623, 108750624, 108750625, 108753175, 108753176, 108753177, 108753178, 108753179, 108766308, 108766309, 108766310, 108766311, 108768265, 108773362, 108773388, 108773389, 108776596, 108776597, 108776598, 108776599, 108784862, 108784865, 108784866, 108784867, 108784868, 108786548, 108786549, 108786550, 108786551, 108786552, 108786553, 108798264, 108799126, 108799128, 108799129, 108799130, 108799131, 108799132, 108799133, 108806749, 108806750, 108806751, 108806752, 108817691, 108817693, 108817694, 108823304, 108826291, 108826292, 108826293, 108826294, 108826295, 108826296, 108832044, 108835017, 108835018, 108835019, 108835020, 108835100, 108835102, 108835103, 108877191, 108877192, 108886196, 108886197, 108886204, 108886205, 108886206, 108886207, 108886208, 108886209, 108891278, 108892971, 108896990, 108909639, 108909640, 108914098, 108914100, 108917065, 108917882, 108917883, 108917891, 108917892, 108917893, 108917894, 108926822, 108926823, 108927498, 108927506, 108927508, 108927509, 108927511, 108938126, 108938128, 108938130, 108952802, 108952803, 108952804, 108952805, 108954128, 108957452, 108957453, 108957454, 108957455, 108959400, 108959401, 108959402, 108959403, 108959404, 108959405, 108961409, 108961410, 108961411, 108961412, 108965179, 108965180, 108965181, 108965182, 108965183, 108970447, 108970448, 108970449, 108970450, 108970451, 108970452, 108970455, 108970456, 108970457, 108973043, 109042912, 109042913, 109042914, 109042915, 109042916, 109042917, 109042918, 109042919, 109042920, 109042921, 109042922, 109042923; c) UDI/DI 00887761968455, Serial Numbers: 108592036, 108611920, 108619324, 108648397, 108648399, 108653145, 108664152, 108665620, 108671478, 108671479, 108671488, 108757703, 108778413, 108784863, 108832046, 108832147, 108835101, 108892991, 108896993, 108896994, 108917886, 108917889, 108933948, 108933952, 108933958, 108941393, 108941394, 108941395, 108941397, 108949663, 108949664, 108949665, 108949666, 108949667, 108949668, 108966760, 108966761, 108966762, 108966763, 108966764, 108966765, 108966766, 108966767, 108973046, 108973047, 108973048, 108973049, 108978929, 108982386, 108982387
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.