AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers.
Published September 12, 2012 · Recall initiated: August 9, 2012
Summary
FDA records medical device Class II recall Z-2321-2012, published September 12, 2012 and initiated August 9, 2012. Recalling firm: AmeriWater Inc; Status: Terminated; Reported quantity: 80 Units; Lot: Model Number MROS, Catalog #00MROS20, 00MROS21 & Model Number MRO1, Catalog #00MRO120, and 00MRO121; Model Number(s): MROS and MRO1; Serial #'s 31000 to 31079. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- AmeriWater Inc
- Distributor
- Worldwide distribution: USA (nationwide) including states of: AL, AR, CO, CT, FL, GA, IL, IN, KS, MA, MD, MO, MS, NC, NY, OH, OK, SC, TN, TX and VA; and country of: Canada.
- Lot
- Model Number MROS, Catalog #00MROS20, 00MROS21 & Model Number MRO1, Catalog #00MRO120, and 00MRO121; Model Number(s): MROS and MRO1; Serial #'s 31000 to 31079.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD GeneOhm VanR Assay , Catalog # 441250, 5 box tests labeled in part***GeneOhm Sciences Canada, Inc., 2555September 12, 2012
- FDASiemens Mobilett Mira. Mobile X-Ray System.September 12, 2012
- FDABD GeneOhm Cdiff Assay , Catalog # 441401, 200 box tests 2-3 labeled in part***GeneOhm Sciences Canada, Inc.,September 12, 2012
- FDAVersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will beSeptember 12, 2012
- FDABD GeneOhm Cdiff Assay , Catalog #441400 200, box tests 1-3 labeled in part***BD Diagnostics, 2555 boul, duSeptember 12, 2012
- FDABrand name: RayStation. Generic name: Treatment Planning System. Software version and build numbers areSeptember 12, 2012
Questions about this recall
Was AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers. recalled?
Yes. FDA published recall Z-2321-2012 on September 12, 2012. Reported hazard: The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.
What should I do if I have this product?
Voluntary: Firm initiated · Letter