Infant Warmer System (IWS)
Published June 21, 2017 · Recall initiated: May 5, 2017
Summary
FDA records medical device Class II recall Z-2323-2017, published June 21, 2017 and initiated May 5, 2017. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 34986; Lot: Medical device listing number: D103053, D103054, D103056, D103057, D183529. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - USA (nationwide) Distribution.
- Lot
- Medical device listing number: D103053, D103054, D103056, D103057, D183529
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAPhilips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, ProductJune 21, 2017
- FDAVideo CytoscopesJune 21, 2017
Questions about this recall
Was Infant Warmer System (IWS) recalled?
Yes. FDA published recall Z-2323-2017 on June 21, 2017. Reported hazard: GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail