Infant Warmer System (IWS)

Published June 21, 2017 · Recall initiated: May 5, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution - USA (nationwide) Distribution.
Lot
Medical device listing number: D103053, D103054, D103056, D103057, D183529
Risk classification
Class II
Area (FDA)
device
Country
United States