ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C
Published July 17, 2024 · Recall initiated: May 31, 2024
Summary
FDA records medical device Class II recall Z-2324-2024, published July 17, 2024 and initiated May 31, 2024. Recalling firm: Zoll Medical Corporation; Status: Ongoing; Reported quantity: 14885; Lot: GTIN: 00812394021239 00812394021277 00812394021284 00812394021413 00812394022779 00847946038182 00847946038199 00847946038243 00847946038250 00847946038304 00847946038328 00847946038335 00847946038342 00847946038359 00847946038380 00847946038397 00847946038403 00847946038410 00847946038434 00847946038441 00847946038496 00847946038519 00847946038533 00847946038571 00847946038588 00847946038663 00847946038816 00847946038830 00847946039028 00847946039035 00847946039066 00847946039127 Serial Number Range: D00000276194-D00000343449. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zoll Medical Corporation
- Distributor
- Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR
- Lot
- GTIN: 00812394021239 00812394021277 00812394021284 00812394021413 00812394022779 00847946038182 00847946038199 00847946038243 00847946038250 00847946038304 00847946038328 00847946038335 00847946038342 00847946038359 00847946038380 00847946038397 00847946038403 00847946038410 00847946038434 00847946038441 00847946038496 00847946038519 00847946038533 00847946038571 00847946038588 00847946038663 00847946038816 00847946038830 00847946039028 00847946039035 00847946039066 00847946039127 Serial Number Range: D00000276194-D00000343449
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAWelch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The ElectrocardiographJuly 17, 2024
- FDAKosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnosticJuly 17, 2024
- FDANon-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examinersJuly 17, 2024
- FDADecompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitionerJuly 17, 2024
- FDA9006767 Astra EV 4.8mm Milled Titanium Abutment Milled abutments are too be used in conjunction with anJuly 17, 2024
- FDAPresource kits and trays, labeled as: 1) ANGIOGRAPHY PACK, Catalog Number SAN13BS605; 2) ANGIOGRAPHY PACK,July 17, 2024
Questions about this recall
Was ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C recalled?
Yes. FDA published recall Z-2324-2024 on July 17, 2024. Reported hazard: G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail