S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors.

Published August 12, 2015 · Recall initiated: July 7, 2015

Summary

FDA records medical device Class II recall Z-2326-2015, published August 12, 2015 and initiated July 7, 2015. Recalling firm: Aesculap, Inc.; Status: Terminated; Reported quantity: 10; Lot: product: FW103R, all Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aesculap, Inc.
Distributor
Distributed US (nationwide) and to Canada.
Lot
product: FW103R, all Lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors. recalled?

Yes. FDA published recall Z-2326-2015 on August 12, 2015. Reported hazard: A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

What should I do if I have this product?

Voluntary: Firm initiated · Letter