da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Published July 11, 2018 · Recall initiated: March 7, 2018
Summary
FDA records medical device Class II recall Z-2341-2018, published July 11, 2018 and initiated March 7, 2018. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 108 ACJ's on 35 surgical systems; Lot: System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286; System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298, P117154308; System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257, P117154305; System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289, P117154290; System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306, P117154307; System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277, P117154296; System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037; System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294, P117154302; System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078; System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083; System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300; System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304; System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093; System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990; System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058 System S/N SL0038, ACJ S/N's P117241971, P117241993; System S/N SL0039, ACJ S/N's P117194013, P117194024, P117194031, P117194032; System S/N SL0040, ACJ S/N's P117194046, P117194061, P117194088, P117194094; System S/N SL0041, ACJ S/N's P117194027, P117194040, P117194043, P117194049; System S/N SL0042, ACJ S/N's P117154288, P117194015, P117194052, P117194090; System S/N SL0043, ACJ S/N P117241988; System S/N SL0044, ACJ S/N's P117194007, P117194085, P117241978; System S/N SL0045, ACJ S/N's P117241965, P117241970, P117241982, P117241984; System S/N SL0046, ACJ S/N's P117194014, P117241964, P117241976, P117241980; System S/N SL0047, ACJ S/N's P117194020, P117194048, P117194067, P117194074 System S/N SL0048, ACJ S/N's P117194004, P117194072, P117200171; System S/N SL0052, ACJ S/N's P117241977, P117241979, P117241986, P117241991; System S/N SL0053, ACJ S/N P117241966; System S/N SL0055, ACJ S/N P117283217; System S/N SL0056, ACJ S/N P117283216; System S/N SL0060, ACJ S/N P117241972; System S/N SL0063, ACJ S/N P117241968; System S/N SL0070, ACJ S/N's P117194028, P117194092; System S/N SL0071, ACJ S/N's P117194018, P117283218; and System S/N SL0073. ACJ S/N P117283214. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- US Nationwide AR, CA, FL, GA, IL, IA, MN, MS, NE, OK, TN, and TX. Foreign distribution was made to Australia, Austria, Finland, France, Germany, Italy, Netherlands, Spain, and United Kingdom.
- Lot
- System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286; System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298, P117154308; System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257, P117154305; System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289, P117154290; System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306, P117154307; System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277, P117154296; System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037; System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294, P117154302; System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078; System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083; System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300; System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304; System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093; System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990; System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058 System S/N SL0038, ACJ S/N's P117241971, P117241993; System S/N SL0039, ACJ S/N's P117194013, P117194024, P117194031, P117194032; System S/N SL0040, ACJ S/N's P117194046, P117194061, P117194088, P117194094; System S/N SL0041, ACJ S/N's P117194027, P117194040, P117194043, P117194049; System S/N SL0042, ACJ S/N's P117154288, P117194015, P117194052, P117194090; System S/N SL0043, ACJ S/N P117241988; System S/N SL0044, ACJ S/N's P117194007, P117194085, P117241978; System S/N SL0045, ACJ S/N's P117241965, P117241970, P117241982, P117241984; System S/N SL0046, ACJ S/N's P117194014, P117241964, P117241976, P117241980; System S/N SL0047, ACJ S/N's P117194020, P117194048, P117194067, P117194074 System S/N SL0048, ACJ S/N's P117194004, P117194072, P117200171; System S/N SL0052, ACJ S/N's P117241977, P117241979, P117241986, P117241991; System S/N SL0053, ACJ S/N P117241966; System S/N SL0055, ACJ S/N P117283217; System S/N SL0056, ACJ S/N P117283216; System S/N SL0060, ACJ S/N P117241972; System S/N SL0063, ACJ S/N P117241968; System S/N SL0070, ACJ S/N's P117194028, P117194092; System S/N SL0071, ACJ S/N's P117194018, P117283218; and System S/N SL0073. ACJ S/N P117283214.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDAGE Healthcare CARESCAPE Monitor B650July 11, 2018
- FDASafety Sideport Knife 1.0mm 45®, REF 378210 Product Usage: Indicated to make stab incisions during ophthalmicJuly 11, 2018
Questions about this recall
Was da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. recalled?
Yes. FDA published recall Z-2341-2018 on July 11, 2018. Reported hazard: Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
What should I do if I have this product?
Voluntary: Firm initiated · Letter