microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
Published July 11, 2018 · Recall initiated: May 10, 2018
Summary
FDA records medical device Class II recall Z-2351-2018, published July 11, 2018 and initiated May 10, 2018. Recalling firm: FHC, Inc.; Status: Terminated; Reported quantity: 2952 total; Lot: UDI: 00873263002162. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- FHC, Inc.
- Distributor
- The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
- Lot
- UDI: 00873263002162
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAmicroTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide aNovember 20, 2024
- FDAFHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulateJuly 15, 2020
- FDAmicroTargeting" STar" Drive System (Manual), Product Number ST-DS-MAJuly 11, 2018
- FDAZimmer® M/L Taper Hip Prosthesis Plasma Sprayed, Size 4 Standard Offset; 00-7711-004-10July 11, 2018
- FDASafety Sideport Knife 1.0mm 45®, REF 378210 Product Usage: Indicated to make stab incisions during ophthalmicJuly 11, 2018
- FDAZimmer® M/L Taper Hip Prosthesis With Kinectiv® Technology Cementless, Size 6; 00-7713-006-00July 11, 2018
Questions about this recall
Was microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA recalled?
Yes. FDA published recall Z-2351-2018 on July 11, 2018. Reported hazard: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter