QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model/Catalog Number: 900370 Software Version: 1.0.1
Published July 24, 2024 · Recall initiated: June 3, 2024
Summary
FDA records medical device Class II recall Z-2353-2024, published July 24, 2024 and initiated June 3, 2024. Recalling firm: Qiagen Sciences LLC; Status: Ongoing; Reported quantity: 419 units; Lot: GTIN 04053228039129 ALL serial numbers shipped to date. SN 30000-30506 is the impacted range (later serial numbers and units in QIAGEN warehouses will include supplemental instructions for operating the heater in protocols requiring incubation temperatures of 40C or lower prior to shipment). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Heater shaker module does not perform heating if the temperature is set below 40®C for static incubation steps when running IVD protocol QIAamp® DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Qiagen Sciences LLC
- Distributor
- Nationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB, ID, IN, IT, JO, KR, KZ, LU, ML, MY, NL, NO, PH, PL, PY, SA, SE, SG, SK, SO, VN J, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI
- Lot
- GTIN 04053228039129 ALL serial numbers shipped to date. SN 30000-30506 is the impacted range (later serial numbers and units in QIAGEN warehouses will include supplemental instructions for operating the heater in protocols requiring incubation temperatures of 40C or lower prior to shipment)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pumpJuly 24, 2024
- FDANitro Glide Knee Walker-offers users recovering from a below-the-knee injury the ability to remain mobile andJuly 24, 2024
- FDAEZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecularApril 17, 2024
- FDAQIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223January 12, 2022
- FDAtherascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121June 17, 2020
- FDAQIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument isAugust 28, 2019
Questions about this recall
Was QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model/Catalog Number: 900370 Software Version: 1.0.1 recalled?
Yes. FDA published recall Z-2353-2024 on July 24, 2024. Reported hazard: Heater shaker module does not perform heating if the temperature is set below 40®C for static incubation steps when running IVD protocol QIAamp® DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail