FDAFDA · Food and drugs·Recall number Z-2357-2019

Microdot Xtra blood glucose test strips

Published August 28, 2019 · Recall initiated: July 26, 2018

Summary

FDA records medical device Class II recall Z-2357-2019, published August 28, 2019 and initiated July 26, 2018. Recalling firm: Cambridge Sensors Limited; Status: Ongoing; Reported quantity: 17,784 (USA); Lot: LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cambridge Sensors Limited
Distributor
Illinois
Lot
LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was Microdot Xtra blood glucose test strips recalled?

Yes. FDA published recall Z-2357-2019 on August 28, 2019. Reported hazard: The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail