OUTLOOK IV SET 15DROP W/2 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication. Model/Catalog Number: 354212
Published July 24, 2024 · Recall initiated: June 17, 2024
Summary
FDA records medical device Class II recall Z-2376-2024, published July 24, 2024 and initiated June 17, 2024. Recalling firm: B. Braun Medical, Inc.; Status: Ongoing; Reported quantity: 236208 units; Lot: UDI-DI 04046964293832. All Lots Distributed After 01AUG2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B. Braun Medical, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the country of Canada.
- Lot
- UDI-DI 04046964293832. All Lots Distributed After 01AUG2023.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock,September 25, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV-Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/ 0.2 FILTER- Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat SPACE PUMP IV SET, 127 IN.-Used with an electrically-powered infusion pump for intravenous (IV)July 24, 2024
- FDAVACUETTE® TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring, non-ridgedJuly 24, 2024
- FDAOUTLOOK IV SET 15DROP W/3 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV)July 24, 2024
Questions about this recall
Was OUTLOOK IV SET 15DROP W/2 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication. Model/Catalog Number: 354212 recalled?
Yes. FDA published recall Z-2376-2024 on July 24, 2024. Reported hazard: Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
What should I do if I have this product?
Voluntary: Firm initiated · Letter