Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Published September 26, 2012 · Recall initiated: August 14, 2012
Summary
FDA records medical device Class II recall Z-2383-2012, published September 26, 2012 and initiated August 14, 2012. Recalling firm: Varian Medical Systems, Inc. Oncology Systems; Status: Terminated; Reported quantity: 221 Units; Lot: LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711 ,H181181 ,H294651 ,H294712 ,H181182 ,H294652 ,H294713 ,H181183 ,H294653 ,H294714 ,H181184 ,H294654 ,H294715 ,H181185 ,H294655 ,H294716 ,H181186 ,H294656 ,H294717 ,H181187 ,H294657 ,H294718 ,H181188 ,H294658 ,H294719 ,H181189 ,H294659 ,H294720 ,H181190 ,H294660 ,H294722 ,H181191 ,H294661 ,H294723 ,H181192 ,H294662 ,H294724 ,H182000 ,H294663 ,H294726 ,H182001 ,H294664 ,H294727 ,H182002 ,H294665 ,H294728 ,H182003 ,H294666 ,H294729 ,H182005 ,H294667 ,H294730 ,H294607 ,H294668 ,H294731. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Varian Medical Systems, Inc. Oncology Systems
- Distributor
- Worldwide Distribution.
- Lot
- LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711 ,H181181 ,H294651 ,H294712 ,H181182 ,H294652 ,H294713 ,H181183 ,H294653 ,H294714 ,H181184 ,H294654 ,H294715 ,H181185 ,H294655 ,H294716 ,H181186 ,H294656 ,H294717 ,H181187 ,H294657 ,H294718 ,H181188 ,H294658 ,H294719 ,H181189 ,H294659 ,H294720 ,H181190 ,H294660 ,H294722 ,H181191 ,H294661 ,H294723 ,H181192 ,H294662 ,H294724 ,H182000 ,H294663 ,H294726 ,H182001 ,H294664 ,H294727 ,H182002 ,H294665 ,H294728 ,H182003 ,H294666 ,H294729 ,H182005 ,H294667 ,H294730 ,H294607 ,H294668 ,H294731
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVarian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29,May 22, 2013
- FDAVarian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage:May 1, 2013
- FDAVarian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse ProtonFebruary 27, 2013
- FDAVarian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import ExportDecember 26, 2012
- FDAUltraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tonesSeptember 26, 2012
- FDAPhilips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-UniversalSeptember 26, 2012
Questions about this recall
Was Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. recalled?
Yes. FDA published recall Z-2383-2012 on September 26, 2012. Reported hazard: Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
What should I do if I have this product?
Voluntary: Firm initiated · Letter