Norton Arm Kit, Product Code 711400, (previous product code 031001), Sample Kit S711400

Published September 8, 2021 · Recall initiated: July 23, 2021

Summary

FDA records medical device Class II recall Z-2390-2021, published September 8, 2021 and initiated July 23, 2021. Recalling firm: Aspen Surgical Products, Inc.; Status: Terminated; Reported quantity: 124; Lot: All lots manufactured between July 9, 2019 to July 9, 2021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aspen Surgical Products, Inc.
Distributor
US nationwide distribution. No foreign distribution.
Lot
All lots manufactured between July 9, 2019 to July 9, 2021
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Norton Arm Kit, Product Code 711400, (previous product code 031001), Sample Kit S711400 recalled?

Yes. FDA published recall Z-2390-2021 on September 8, 2021. Reported hazard: Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.