Reliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theatre shoes and other miscellaneous reusable items used in the care of patients.
Published July 18, 2018 · Recall initiated: May 23, 2018
Summary
FDA records medical device Class II recall Z-2392-2018, published July 18, 2018 and initiated May 23, 2018. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 118; Lot: 3632511020 3634311005 3607312011 3615312001 3611012005 3627012001 3626212001 3633112020 3632612001 3601413013 3629612014 3602113013 3605113001 3615912011 3606613001A 3607413001 3612713001 3614913001 3616213001 3608613001 3623313001 3624113001 3626813001 3627313019 3628113001 3628213002 3629113001 3631613011 3635013015 3601514001 3603614001 3604114024 3604914002 3630113020 3613214010 3611814020 3616714023 3630813004 3628214010 3629414002 3630814002 3631614001 3631414016 3632914001 3632414003 3634914022 3604415042 3605415029 3606115033 3604715027 3608215021 3615215026 3613915001 3619415002 3623015006 3623715001 3627115035 3627315001 3627815020 3629215021 3629515001 3633015001 3632215001 3633615001 3634215001 3632315001 3626415043 3601116014 3602516001 3634315001 3604116001 3604616003 3604716001 3604816001 3606916011 3613716013 3618716001 3600616001 3625616008 3630116001 3632616007 3634216002 3633316017 3635416008 3625716001 3635616001 3603217002 3632116001 3635016001 3604517001 3604617001 3605417001 3606117001 3628616001 3605317001 3605117029 3607417001 3611417015 3612217002 3610917001 3611617001 3613617001 3613717001 3619817008 3619117006 3619917001 3619317001 3617717005 3630317038 3634117001 3601118001 3634517031 3606418014 3606518001 3608718001 3609318016 3610118001 3616814019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide Distribution - US Nationwide in the states of AL, AK, CA, CO, CN, DE, FL, GA, IL, IA, KS, KY, LA, MA, MI, NJ, NY, ND, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WI, and WV. Foreign distribution to Canada.
- Lot
- 3632511020 3634311005 3607312011 3615312001 3611012005 3627012001 3626212001 3633112020 3632612001 3601413013 3629612014 3602113013 3605113001 3615912011 3606613001A 3607413001 3612713001 3614913001 3616213001 3608613001 3623313001 3624113001 3626813001 3627313019 3628113001 3628213002 3629113001 3631613011 3635013015 3601514001 3603614001 3604114024 3604914002 3630113020 3613214010 3611814020 3616714023 3630813004 3628214010 3629414002 3630814002 3631614001 3631414016 3632914001 3632414003 3634914022 3604415042 3605415029 3606115033 3604715027 3608215021 3615215026 3613915001 3619415002 3623015006 3623715001 3627115035 3627315001 3627815020 3629215021 3629515001 3633015001 3632215001 3633615001 3634215001 3632315001 3626415043 3601116014 3602516001 3634315001 3604116001 3604616003 3604716001 3604816001 3606916011 3613716013 3618716001 3600616001 3625616008 3630116001 3632616007 3634216002 3633316017 3635416008 3625716001 3635616001 3603217002 3632116001 3635016001 3604517001 3604617001 3605417001 3606117001 3628616001 3605317001 3605117029 3607417001 3611417015 3612217002 3610917001 3611617001 3613617001 3613717001 3619817008 3619117006 3619917001 3619317001 3617717005 3630317038 3634117001 3601118001 3634517031 3606418014 3606518001 3608718001 3609318016 3610118001 3616814019
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAInnova IGS 630/520/530/540 Interventional Fluoroscopic X-ray SystemsJuly 18, 2018
- FDAFemoral Nail, A/R T2 Femur ¿10x320 mm, Product Number 18251032SJuly 18, 2018
Questions about this recall
Was Reliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theatre shoes and other miscellaneous reusable items used in the care of patients. recalled?
Yes. FDA published recall Z-2392-2018 on July 18, 2018. Reported hazard: The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.
What should I do if I have this product?
Voluntary: Firm initiated · Letter