Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
Published July 8, 2020 · Recall initiated: May 20, 2020
Summary
FDA records medical device Class I recall Z-2413-2020, published July 8, 2020 and initiated May 20, 2020. Recalling firm: Arrow International Inc; Status: Terminated; Reported quantity: 2123 total; Lot: Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120W 150201W 150305W 150307W through 150309W 150623W through 150624W 150713W through 150716W 151114W 151209W 151213W 151219W 160503W 160511W through 160515W 160517W 160635W through 160636W 160640W through 160641W 160731W through 160733W 171105W 171117W 171119W through 171120W 171201W 180120W 180301W. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Arrow International Inc
- Distributor
- Worldwide distribution.
- Lot
- Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120W 150201W 150305W 150307W through 150309W 150623W through 150624W 150713W through 150716W 151114W 151209W 151213W 151219W 160503W 160511W through 160515W 160517W 160635W through 160636W 160640W through 160641W 160731W through 160733W 171105W 171117W 171119W through 171120W 171201W 180120W 180301W
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400July 8, 2020
- FDAArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701July 8, 2020
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500July 8, 2020
- FDANeo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible InfantJuly 8, 2020
- FDAEnlightened Dairy-Free Chocolate Peanut Butter Frozen Dessert, one Pint (473 ml). Case GTIN:July 8, 2020
- FDAArrow StimuCath(R) Continuous Nerve Block Kit ASK-05060-CHO1February 26, 2020
Questions about this recall
Was Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J recalled?
Yes. FDA published recall Z-2413-2020 on July 8, 2020. Reported hazard: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
What should I do if I have this product?
Voluntary: Firm initiated · Letter