FDAFDA · Food and drugs·Recall number Z-2417-2020

Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701

Published July 8, 2020 · Recall initiated: May 20, 2020

Summary

FDA records medical device Class I recall Z-2417-2020, published July 8, 2020 and initiated May 20, 2020. Recalling firm: Arrow International Inc; Status: Terminated; Reported quantity: 2123 total; Lot: Serial Number Ranges: AC3 Optimus IABP NA/AJLA IAP-0701 170831F through 170840F 171011F through 171020F 171121F through 171125F 180106F through 180110F 180126F through 180130F 180426F through 180430F 180606F through 180610F 180731F through 180740F 180826F through 180830F 180901F through 180910F 180912F through 180921F 181001F through 181010F 190136F through 190140F 190336F through 190340F 190646F through 190650F 190731F through 190736F 190931F through 190936F 191019F through 191024F 191055F through 191060F 200107F through 200112F 200308F through 200311F. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Arrow International Inc
Distributor
Worldwide distribution.
Lot
Serial Number Ranges: AC3 Optimus IABP NA/AJLA IAP-0701 170831F through 170840F 171011F through 171020F 171121F through 171125F 180106F through 180110F 180126F through 180130F 180426F through 180430F 180606F through 180610F 180731F through 180740F 180826F through 180830F 180901F through 180910F 180912F through 180921F 181001F through 181010F 190136F through 190140F 190336F through 190340F 190646F through 190650F 190731F through 190736F 190931F through 190936F 191019F through 191024F 191055F through 191060F 200107F through 200112F 200308F through 200311F
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701 recalled?

Yes. FDA published recall Z-2417-2020 on July 8, 2020. Reported hazard: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

What should I do if I have this product?

Voluntary: Firm initiated · Letter