Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

Published July 31, 2024 · Recall initiated: June 4, 2024

Summary

FDA records medical device Class II recall Z-2414-2024, published July 31, 2024 and initiated June 4, 2024. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 143 devices; Lot: Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Lot
Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137. recalled?

Yes. FDA published recall Z-2414-2024 on July 31, 2024. Reported hazard: The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter