STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
Published September 26, 2012 · Recall initiated: April 6, 2012
Summary
FDA records medical device Class II recall Z-2426-2012, published September 26, 2012 and initiated April 6, 2012. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 379 tables; Lot: Serial numbers from 0403809001 to 0408612102. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide Distribution - US Nationwide including the states of: AL, AR, CA, CO, CT, FL, HI, ID, IL, IN, IA, KY, LA, MA, MD, ME, MI, MN, MO, MS, NV, NYH, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WI & WV and countries of: Australia, Canada, China, India, Korea, Mexico, Panama and Thailand.
- Lot
- Serial numbers from 0403809001 to 0408612102
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAULTRA Fabric Reinforced Surgical Gown, X-Large, Sterile, Test File. The Kimberly-Clark ULTRA Fabric ReinforcedSeptember 26, 2012
- FDA12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REFSeptember 26, 2012
Questions about this recall
Was STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery). recalled?
Yes. FDA published recall Z-2426-2012 on September 26, 2012. Reported hazard: STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
What should I do if I have this product?
Voluntary: Firm initiated · Letter