Sensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. Equipped by modules, enabling various configurations.
Published July 31, 2024 · Recall initiated: July 3, 2024
Summary
FDA records medical device Class II recall Z-2439-2024, published July 31, 2024 and initiated July 3, 2024. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Ongoing; Reported quantity: 7 systems; Lot: UDI-DI: N/A Serial Numbers: Serial 18023 18022 18020 18021 18019 18025 18017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide distribution in the states of CA, CT, FL, IA, KS, MA, MI, NJ, TX, VA.
- Lot
- UDI-DI: N/A Serial Numbers: Serial 18023 18022 18020 18021 18019 18025 18017
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDASensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to beJuly 31, 2024
- FDASensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post ProcessingJuly 31, 2024
Questions about this recall
Was Sensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. Equipped by modules, enabling various configurations. recalled?
Yes. FDA published recall Z-2439-2024 on July 31, 2024. Reported hazard: If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
What should I do if I have this product?
Voluntary: Firm initiated · Letter