Computed Tomography X-Ray Systems, (Brilliance 6, Brilliance 16, MX8000 Dual v. Exp, GEMINI Dual), Philips Healthcare Systems, Cleveland, OH. The Brilliance 6, 16 and MX8000 Dual v. Exp are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. GEMINI Dual is an imaging system that combines positron emission tomography (PET) and X-ray computed tomography. The GEMINI produces attenuation and non-attenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as X-ray transmission images of these areas. The PET and CT images are registered and displayed in a fused format to provide combined PET and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. GEMINI can provide CT data suitable for use in attenuation correction.

Published October 3, 2012 · Recall initiated: August 21, 2012

Summary

FDA records medical device Class II recall Z-2451-2012, published October 3, 2012 and initiated August 21, 2012. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 112 Units; Lot: System Code #: 728130; Serial #: 8128, 8158 & 8369. System Code #: 728256; Serial #: 3695, 3703, 371 &, 3719. System Code #: 728246; Serial #: 3463, 3508, 3519, 3548, 3552, 3591, 3592, 3593, 3603, 3637, 3648, 3654, 3659, 3661, 3675, 3679, 3687, 3692, 5451, 5515, 5543, 5567, 5583, 5586, 5658, 5664, 5675, 5697, 5707, 5749, 5752, 5774, 5807, 5813, 5815, 5845, 5846, 5852, 5866, 5872, 5910, 5963, 5968, 6028, 6036, 6045, 6071 & 6126. System Code #: 882160; Serial #: 97, 102, 103, 105, 106, 107, 109, 110, 111, 113, 114, 115, 116, 117, 118, 121, 122, 123, 124, 125, 126, 127, 131, 132, 133, 134, 135, 136, 137, 139, 140, 141, 143, 144, 146, 147, 148, 149, 150, 151, 152, 153, 154, 155, 156, 157, 158, 159, 160, 161, 162, , 63, 164, 165, 167, 1065 & 6119. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - USA including AL, AZ, CA, CO, FL, GA, IL, KS, KY, MI, MS, NJ, NM, NV, NY, OH, TX, WA & WI and internationally to: Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, India, Italy, Japan, Jordan, Mexico, Saudi Arabia, South Korea, Spain, Thailand, Turkey & United Kingdom.
Lot
System Code #: 728130; Serial #: 8128, 8158 & 8369. System Code #: 728256; Serial #: 3695, 3703, 371 &, 3719. System Code #: 728246; Serial #: 3463, 3508, 3519, 3548, 3552, 3591, 3592, 3593, 3603, 3637, 3648, 3654, 3659, 3661, 3675, 3679, 3687, 3692, 5451, 5515, 5543, 5567, 5583, 5586, 5658, 5664, 5675, 5697, 5707, 5749, 5752, 5774, 5807, 5813, 5815, 5845, 5846, 5852, 5866, 5872, 5910, 5963, 5968, 6028, 6036, 6045, 6071 & 6126. System Code #: 882160; Serial #: 97, 102, 103, 105, 106, 107, 109, 110, 111, 113, 114, 115, 116, 117, 118, 121, 122, 123, 124, 125, 126, 127, 131, 132, 133, 134, 135, 136, 137, 139, 140, 141, 143, 144, 146, 147, 148, 149, 150, 151, 152, 153, 154, 155, 156, 157, 158, 159, 160, 161, 162, , 63, 164, 165, 167, 1065 & 6119.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Computed Tomography X-Ray Systems, (Brilliance 6, Brilliance 16, MX8000 Dual v. Exp, GEMINI Dual), Philips Healthcare Systems, Cleveland, OH. The Brilliance 6, 16 and MX8000 Dual v. Exp are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. GEMINI Dual is an imaging system that combines positron emission tomography (PET) and X-ray computed tomography. The GEMINI produces attenuation and non-attenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as X-ray transmission images of these areas. The PET and CT images are registered and displayed in a fused format to provide combined PET and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. GEMINI can provide CT data suitable for use in attenuation correction. recalled?

Yes. FDA published recall Z-2451-2012 on October 3, 2012. Reported hazard: A bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.

What should I do if I have this product?

Voluntary: Firm initiated · Letter