Back to search
FDAFDA · Food and drugs·Recall number Z-2456-2026

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Published June 24, 2026 · Recall initiated: March 16, 2026

Summary

FDA records medical device Class II recall Z-2456-2026, published June 24, 2026 and initiated March 16, 2026. Recalling firm: Smith & Nephew, Inc.; Status: Ongoing; Reported quantity: 5; Lot: UDI-DI: 03596010514394. Lot/Batch: 25BM03616. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smith & Nephew, Inc.
Distributor
International distribution to the countries of BE, DE, IT.
Lot
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816 recalled?

Yes. FDA published recall Z-2456-2026 on June 24, 2026. Reported hazard: Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

What should I do if I have this product?

Voluntary: Firm initiated · Letter