Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Published June 24, 2026 · Recall initiated: May 18, 2026
Summary
FDA records medical device Class I recall Z-2457-2026, published June 24, 2026 and initiated May 18, 2026. Recalling firm: Abiomed, Inc.; Status: Ongoing; Reported quantity: 10 units; Lot: GTIN: 00813502012279. Product Codes (Serial Number): 1000080 (677223, 672986, 673252, 644591, 645428, 644314, 613525), 1000851 (636173), 0048-0014 (499250), 0048-0024-JP (661216). Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abiomed, Inc.
- Distributor
- US distribution to GA, IA, NE, PA, TX, WA, WI. International distribution to Czech Republic, Norway, and Japan
- Lot
- GTIN: 00813502012279. Product Codes (Serial Number): 1000080 (677223, 672986, 673252, 644591, 645428, 644314, 613525), 1000851 (636173), 0048-0014 (499250), 0048-0024-JP (661216).
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAImpella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, aNovember 4, 2026
- FDAImpella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, aNovember 4, 2026
- FDAAutomated Impella Controller, Packaged. Primary user control interface for the Impella Catheter,November 4, 2026
- FDAAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1.September 30, 2026
- FDATRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816June 24, 2026
- FDARitmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C;June 24, 2026
Questions about this recall
Was Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP. recalled?
Yes. FDA published recall Z-2457-2026 on June 24, 2026. Reported hazard: Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
What should I do if I have this product?
Voluntary: Firm initiated · Letter