SIGNA Pioneer, with affected software versions: PX25.0 to PX25.5, PX26.0, PX26.1, PX28.0 to PX28.4, PX29.0, PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System

Published August 14, 2024 · Recall initiated: June 14, 2024

Summary

FDA records medical device Class II recall Z-2460-2024, published August 14, 2024 and initiated June 14, 2024. Recalling firm: GE Medical Systems, LLC; Status: Ongoing; Reported quantity: 956 units; Lot: System ID: 520744MR, 714578MR1, RADNETMR1727, RADNETMR652, 951486MR2, RADNETMR651, 831RMGMR1, RADNETMR647, RADNETMR661, 720442MR, 303508MR, 860584MR3T, 203694MR3T, 860678MR3T, 860545MR3T, 860536MR3T, 203327MR3T, 860871MR3T, 203949MR3T, 305933MR1, 954489MR2, 954492MR1, 407ODMR3T, 863577MR, 863607MR, 386385MR1, 850696MR3T, 954423MR2, RADNETMR1700, 850877PNR, 706PIONEER, 847459MR, 815215FDMR, 217223BHMR3T, 217QMG3TMR, 574SBOPIONEER, 318769MR2, 504LCMCHMR, 985WCMR3, 508872PMR, 508998MR, 413SHMR3T, 508775MR1, 781331MR, 508756PMR, 410266MR5, RADNETMR1719, RADNETMR1713, RADNETMR1708, 443481MR, RADNETMR1720, 301279MR4, RADNETMR1696, RADNETMR1724, RADNETMR1695, 828250MRW, 704362PIO, 704384PIO, 704333PIO, 919471DDIMR, 910483CI3T, 336397P3T, 856532MR, 908735PMR, 201796MR2, 973301MR3, 732462MRPNR, RADNETMR1711, 856231MR, SDMIADPMR1, SDMINWPMR1, SDMIBDPMR2, SDMICR3TPMR1, 718596MR, 718245MR, 631CHSPNR, 631858MR, RADNETMR1702, 315792MR, 516NYCB3T1, 516NYCB3T2, RADNETMR1704, 716278P3T, 516766MR3T, 631751MR, 518926MR3T, 518880MR, 516340MR3T, 315SDMG3T, 716656P3T, RADNETMR1707, 610402MR3T, 4603875-MDF035-1-1-1, 864225MR3T, 843681BL3T, 843797ARTMR, 8435563T, 843234MR3T, 843777PL3T, 843777PIO3T, 864255MMR, 803936LMCMR2, SPORTSMED3T, 423553MRD, 972747EAMR, 361275PIONMR, 469648EOT3T, 214320DFWD3T, 214954PCDMR, 469902EFM3T, 817741ENFW, 817592ESFW, 817433PHMR1, 830HCMPMR, 972805ENFMR1, 972495TSGMR2, 713512MHMR1, MHNEMR1, 972258HAMR1, 281HMI750W, RADNETMR678, 832638SCMR1, 512APIRRMR, 210MSNWPMR, 817424TSSL3T, 801ENVBMR3, 801377MR6, 757578GE3T, 262289MR1, LXTPMSW1NET, LXTT15NET, LXTT43NET, 307733MR, MR5314AR, MR3048AR, MR6508AR, MR6802AR, MR6623AR, MR52800461, 83203735240, 83183534632, 83183547990, 83141851264, 83183543071, 83144557293, 83178104799, 834MR000002, 83183539988, 834MR000005, 834MR000006, 83132425984, 834MR000050, 834MR000021, 834MR000015, 83155846100, 834MR000013MM4, 83210367946, 83178782037, 83201375088, 83152240359, 833MR000024, 833MR000024MM4, 83199160995, 83170864668, 83153235400, 083027603871918, 083027600728323, 083027600259223, 083027600101122, 083027607448619, BY4197MR01, 190207MR01, 110009MR01, 150019MR01, 110059MR01, MRUA2341, MRUA2246, MRUA0897, MRUA0751, MRUA0592, MRUA1533, MRUA2387, MRUA1058, MRUA2091, MRUA1977, MRUA1479, MRUA0790, MRUA2067, MRUA1053A, MRUA1135, MRUA2262, MRUA2182, MRUA1761, MRUA1746, MRUA0774, MRUA1294, MRUA2012, MRUA2303, MRUA2342, MRUA1595, MRUA1721, BG4605MR01, 403253MR1, 250598MR1, MR2470CL, MRUA1674, 082427110150, 082427110108, 082427110148, 082427040339, 082427040219, 082427040202, 082427040245, 082427220133, 082427230246, 082427090424, 082427010211, 082427010142, 082427060114, 082427100458, 082427100420, 082427290066, 082427030188, 082427160345, 082427160277, 082427070336, 082427070237, 082427090336, 082427240106, 082427040260, 082427040234, 082427040211, 082427150102, 082427020100, 082427080114, 082427060086, 082427140324, 082427050050, 082427050048, 082427100387, 082427200211, 082427090368, 082427070341, 082427260011, 082427250078, 082427080090, 082427030161, 082427030184, 082427140204, 082427030154, 082427100416, 082427100419, 082427090454, 082427100357, 082427010204, 082427100422, 082427250087, 082427030193, 082427090397, 082427090346, 082427220128, 082427020095, 082427100365, 082427030147, 082427010161, 082427290065, 082427290072, 082427160194, 082427160338, 082427160306, 082427160264, 082427160395, 082427160398, 082427160363, 082427160294, 082427180086, 082427240107, 082427240075, 082427240089, 082427170033, 082427170047, 082427210168, 082427210200, 082427090369, 082427070269, 082427070328, 082427070333, 082427070334, 082427070221, 082427070326, 082427070250, 082427070291, 082427070228, 082427070192, 082427070302, 082427070259, 082427070220, 082427010206, 082427010167, 082427010213, 082427090377, 082427090374, 082427090382, 082427090360, 082427090383, 082427090340, 082427090355, 082427090332, 082427090339, 082427090353, 082427090399, 082427090348, 082427090412, 082427210206, 082427190193, 082427. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Medical Systems, LLC
Distributor
Worldwide distribution.
Lot
System ID: 520744MR, 714578MR1, RADNETMR1727, RADNETMR652, 951486MR2, RADNETMR651, 831RMGMR1, RADNETMR647, RADNETMR661, 720442MR, 303508MR, 860584MR3T, 203694MR3T, 860678MR3T, 860545MR3T, 860536MR3T, 203327MR3T, 860871MR3T, 203949MR3T, 305933MR1, 954489MR2, 954492MR1, 407ODMR3T, 863577MR, 863607MR, 386385MR1, 850696MR3T, 954423MR2, RADNETMR1700, 850877PNR, 706PIONEER, 847459MR, 815215FDMR, 217223BHMR3T, 217QMG3TMR, 574SBOPIONEER, 318769MR2, 504LCMCHMR, 985WCMR3, 508872PMR, 508998MR, 413SHMR3T, 508775MR1, 781331MR, 508756PMR, 410266MR5, RADNETMR1719, RADNETMR1713, RADNETMR1708, 443481MR, RADNETMR1720, 301279MR4, RADNETMR1696, RADNETMR1724, RADNETMR1695, 828250MRW, 704362PIO, 704384PIO, 704333PIO, 919471DDIMR, 910483CI3T, 336397P3T, 856532MR, 908735PMR, 201796MR2, 973301MR3, 732462MRPNR, RADNETMR1711, 856231MR, SDMIADPMR1, SDMINWPMR1, SDMIBDPMR2, SDMICR3TPMR1, 718596MR, 718245MR, 631CHSPNR, 631858MR, RADNETMR1702, 315792MR, 516NYCB3T1, 516NYCB3T2, RADNETMR1704, 716278P3T, 516766MR3T, 631751MR, 518926MR3T, 518880MR, 516340MR3T, 315SDMG3T, 716656P3T, RADNETMR1707, 610402MR3T, 4603875-MDF035-1-1-1, 864225MR3T, 843681BL3T, 843797ARTMR, 8435563T, 843234MR3T, 843777PL3T, 843777PIO3T, 864255MMR, 803936LMCMR2, SPORTSMED3T, 423553MRD, 972747EAMR, 361275PIONMR, 469648EOT3T, 214320DFWD3T, 214954PCDMR, 469902EFM3T, 817741ENFW, 817592ESFW, 817433PHMR1, 830HCMPMR, 972805ENFMR1, 972495TSGMR2, 713512MHMR1, MHNEMR1, 972258HAMR1, 281HMI750W, RADNETMR678, 832638SCMR1, 512APIRRMR, 210MSNWPMR, 817424TSSL3T, 801ENVBMR3, 801377MR6, 757578GE3T, 262289MR1, LXTPMSW1NET, LXTT15NET, LXTT43NET, 307733MR, MR5314AR, MR3048AR, MR6508AR, MR6802AR, MR6623AR, MR52800461, 83203735240, 83183534632, 83183547990, 83141851264, 83183543071, 83144557293, 83178104799, 834MR000002, 83183539988, 834MR000005, 834MR000006, 83132425984, 834MR000050, 834MR000021, 834MR000015, 83155846100, 834MR000013MM4, 83210367946, 83178782037, 83201375088, 83152240359, 833MR000024, 833MR000024MM4, 83199160995, 83170864668, 83153235400, 083027603871918, 083027600728323, 083027600259223, 083027600101122, 083027607448619, BY4197MR01, 190207MR01, 110009MR01, 150019MR01, 110059MR01, MRUA2341, MRUA2246, MRUA0897, MRUA0751, MRUA0592, MRUA1533, MRUA2387, MRUA1058, MRUA2091, MRUA1977, MRUA1479, MRUA0790, MRUA2067, MRUA1053A, MRUA1135, MRUA2262, MRUA2182, MRUA1761, MRUA1746, MRUA0774, MRUA1294, MRUA2012, MRUA2303, MRUA2342, MRUA1595, MRUA1721, BG4605MR01, 403253MR1, 250598MR1, MR2470CL, MRUA1674, 082427110150, 082427110108, 082427110148, 082427040339, 082427040219, 082427040202, 082427040245, 082427220133, 082427230246, 082427090424, 082427010211, 082427010142, 082427060114, 082427100458, 082427100420, 082427290066, 082427030188, 082427160345, 082427160277, 082427070336, 082427070237, 082427090336, 082427240106, 082427040260, 082427040234, 082427040211, 082427150102, 082427020100, 082427080114, 082427060086, 082427140324, 082427050050, 082427050048, 082427100387, 082427200211, 082427090368, 082427070341, 082427260011, 082427250078, 082427080090, 082427030161, 082427030184, 082427140204, 082427030154, 082427100416, 082427100419, 082427090454, 082427100357, 082427010204, 082427100422, 082427250087, 082427030193, 082427090397, 082427090346, 082427220128, 082427020095, 082427100365, 082427030147, 082427010161, 082427290065, 082427290072, 082427160194, 082427160338, 082427160306, 082427160264, 082427160395, 082427160398, 082427160363, 082427160294, 082427180086, 082427240107, 082427240075, 082427240089, 082427170033, 082427170047, 082427210168, 082427210200, 082427090369, 082427070269, 082427070328, 082427070333, 082427070334, 082427070221, 082427070326, 082427070250, 082427070291, 082427070228, 082427070192, 082427070302, 082427070259, 082427070220, 082427010206, 082427010167, 082427010213, 082427090377, 082427090374, 082427090382, 082427090360, 082427090383, 082427090340, 082427090355, 082427090332, 082427090339, 082427090353, 082427090399, 082427090348, 082427090412, 082427210206, 082427190193, 082427
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SIGNA Pioneer, with affected software versions: PX25.0 to PX25.5, PX26.0, PX26.1, PX28.0 to PX28.4, PX29.0, PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System recalled?

Yes. FDA published recall Z-2460-2024 on August 14, 2024. Reported hazard: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

What should I do if I have this product?

Voluntary: Firm initiated · Letter