Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Published August 14, 2024 · Recall initiated: June 14, 2024
Summary
FDA records medical device Class II recall Z-2464-2024, published August 14, 2024 and initiated June 14, 2024. Recalling firm: GE Medical Systems, LLC; Status: Ongoing; Reported quantity: 484 units; Lot: System ID: 205592450W, 205592750, BMDACCMR1, 480425MR6, 623SMCSMR, 602839MR1, 602RITMR, 480768MR2, RADNETMR644, UCB3TMR, 3106573TMR, 650PH3T, 760SMCBMR, 510537MR750, 559325CIIMR3, 858822HDMR3, 858822HDMR1, 209225VC750, ASLW750, 415723CHMR3, 415723SCHMR2, 415493VMR2, 415723CHMR2, 650780MR750, MDF013-1-2OS0CA8W, RADNETMR645, 916LST3TMR, 858939MR5, 858249HLIMR1, MDU009-GE13+4242063+5P1, 415UCBHMR2, 415UCBHMR1, 415OI3TMR, 415L3TMR1, 415L3TMR2, 408358SHMR13, 805687MR1, 415723SHMR19, 415723SHMR21, 415723SHMR16, INSIGHTMR3228, 303699MR2, 720321MR, 202476MR3, 302744MR750, DUPONT3T, 386FMR3T, 561495MR1, 954772MR1, 904202MR9, 904SM3T, 863688MR, 727698MR17, MUNROEMR, AUI3T, 850494450W, 850RA450W1, 813972VAMR, 706UGA3T2, 515222MR1, 847618MR1, 312355DVMR, 312996MR3T, 708349DVMR, 309683MNMR750, 217223BHMR, 630CDH3T, 765494MR750, 617726MR1A2, 301295MR3DV, 301496MR3T5MR, 301295CRADA, 301295MR3T2, 301295MR4, 301496MR3T6MR, 240296MR3T, 202476MR4, 207945MR750, 7347633TMRFIX, UM750MR, 810235DVMR, 6162673T750, 616494DVMR, 2486613TMR, MR129037, MR129035, MR129034, MR129030, 816698CP3T, 816932DVMR, 704384DVMR, 910907WAMR1, 6036503T, 732262MRGS, 201634MRW, 7168453T, CORT450W, 607CORNELL3T, CBCIMR750, MSK65MR2, MDU018-1-2Q5XBKV2, MDU018-1-2Q5XBKZP, HSSMAGNETC, 585785MR750, 614722MR3, 405752I3T, 918491MR2, 541CMRIMR, 503331MR1, 570271MR6, 412623MR1, 412647PUHMR1, 615MTMR1, 4192142MR91, 713792AMR17, 713566SMITH2, 713792CMR1, 281477MR2, 713441MR7, 210LAFB750, 210614MR1, 281274MR1, 281276MR750, 903525MR750, 801377MR3, 801268MR2, 434UVAMR, 703573DV4, 703391MR2, 703805MR3DV, 540MIF3T750, 7579533T, 804675VA3T, 540665MR3T, 425750PMR, 206762VAMR, 206320CH450W, 206CH750MR, 509473DMR3T, 715838MR3T, 608417MR1, 608890MR750, 608HERIMR, 608262RP4, 414805MR4, MRUA0136, MRUA0343, MRWB0843, 1063THDX, 0910272071, 0910272072, 0910274032, 83096322711, 83160776416, 110028MR01, 2706164, MRUA0272, MRHM0942, MRUA0447, MRWB0558, MR13905YR5, 2875893, MRHM1586, MR13708YR3, 2744287, 2655437, 902473MR2, 905522MR3T, 519685MR3T, 514848MR1, 416595MR1, 604875MR750, 082427110049, 082427040163, 082427040199, 082427040187, 082427040155, 082427040153, 082427040154, 082427040143, 082427040123, 082427040151, 082427040141, 082427040224, 082427040157, 082427220040, 082427040148, 082427140107, 082427010056, 082427080068, 082427100184, 082427160114, 082427250070, 082427140176, 082427310077, 082427160118, 082427100311, 082427110080, 082427090169, 082427060042, 082427240082, 082427210226, 082427150073, 082427150075, 082427150058, 082427150077, 082427160128, 082427290031, 082427160151, 082427210136, 082427070075, 082427070176, 082427070212, 082427070079, 082427010073, 082427090241, 082427090298, 082427090269, 082427210068, 082427190118, 082427120141, 082427120107, 082427120132, 082427120145, 082427120101, 082427120092, 082427200076, 082427200110, 082427020075, 082427020039, 082427030127, 082427030100, 082427030076, 082427210116, 082427230051, 082427140106, 082427140076, 082427130089, 082427130078, 082427030092, 082427310062, 082427130079, 082427130080, 082427130075, 082427130072, 082427130096, 082427290023, 082427230106, 082427040205, 082427040249, 082427040208, 082427040180, 082427040177, 082427040189, 082427040174, 082427040166, 082427040172, 082427040165, 082427040137, 082427040133, 082427040132, 082427040127, 082427040120, 082427040119, 082427040116, 082427120150, 082427180027, 082427200058, 082427150070, 082427150047, 082427110084, 082427020042, 082427010081, 082427050037, 082427050034, 082427050028, 082427110058, 082427090283, 082427160129, 082427160088, 082427090201, 082427120097, 082427140130, 082427180031, 082427180029, 082427180022, 082427210121, 082427030094, 082427150090, 082427100215, 082427170029, 082427160239, 082427140098, 082427100223, 082427100339, 082427100335, 082427150082, 082427140122, 082427210081, 082427310057, 082427120129, 082427090308, 082427090311, 082427090261, 082427090148, 082427020056, 082427190111, 082427100319, 082427090256, 082427060106, 082427260005, 0824270800. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide distribution.
- Lot
- System ID: 205592450W, 205592750, BMDACCMR1, 480425MR6, 623SMCSMR, 602839MR1, 602RITMR, 480768MR2, RADNETMR644, UCB3TMR, 3106573TMR, 650PH3T, 760SMCBMR, 510537MR750, 559325CIIMR3, 858822HDMR3, 858822HDMR1, 209225VC750, ASLW750, 415723CHMR3, 415723SCHMR2, 415493VMR2, 415723CHMR2, 650780MR750, MDF013-1-2OS0CA8W, RADNETMR645, 916LST3TMR, 858939MR5, 858249HLIMR1, MDU009-GE13+4242063+5P1, 415UCBHMR2, 415UCBHMR1, 415OI3TMR, 415L3TMR1, 415L3TMR2, 408358SHMR13, 805687MR1, 415723SHMR19, 415723SHMR21, 415723SHMR16, INSIGHTMR3228, 303699MR2, 720321MR, 202476MR3, 302744MR750, DUPONT3T, 386FMR3T, 561495MR1, 954772MR1, 904202MR9, 904SM3T, 863688MR, 727698MR17, MUNROEMR, AUI3T, 850494450W, 850RA450W1, 813972VAMR, 706UGA3T2, 515222MR1, 847618MR1, 312355DVMR, 312996MR3T, 708349DVMR, 309683MNMR750, 217223BHMR, 630CDH3T, 765494MR750, 617726MR1A2, 301295MR3DV, 301496MR3T5MR, 301295CRADA, 301295MR3T2, 301295MR4, 301496MR3T6MR, 240296MR3T, 202476MR4, 207945MR750, 7347633TMRFIX, UM750MR, 810235DVMR, 6162673T750, 616494DVMR, 2486613TMR, MR129037, MR129035, MR129034, MR129030, 816698CP3T, 816932DVMR, 704384DVMR, 910907WAMR1, 6036503T, 732262MRGS, 201634MRW, 7168453T, CORT450W, 607CORNELL3T, CBCIMR750, MSK65MR2, MDU018-1-2Q5XBKV2, MDU018-1-2Q5XBKZP, HSSMAGNETC, 585785MR750, 614722MR3, 405752I3T, 918491MR2, 541CMRIMR, 503331MR1, 570271MR6, 412623MR1, 412647PUHMR1, 615MTMR1, 4192142MR91, 713792AMR17, 713566SMITH2, 713792CMR1, 281477MR2, 713441MR7, 210LAFB750, 210614MR1, 281274MR1, 281276MR750, 903525MR750, 801377MR3, 801268MR2, 434UVAMR, 703573DV4, 703391MR2, 703805MR3DV, 540MIF3T750, 7579533T, 804675VA3T, 540665MR3T, 425750PMR, 206762VAMR, 206320CH450W, 206CH750MR, 509473DMR3T, 715838MR3T, 608417MR1, 608890MR750, 608HERIMR, 608262RP4, 414805MR4, MRUA0136, MRUA0343, MRWB0843, 1063THDX, 0910272071, 0910272072, 0910274032, 83096322711, 83160776416, 110028MR01, 2706164, MRUA0272, MRHM0942, MRUA0447, MRWB0558, MR13905YR5, 2875893, MRHM1586, MR13708YR3, 2744287, 2655437, 902473MR2, 905522MR3T, 519685MR3T, 514848MR1, 416595MR1, 604875MR750, 082427110049, 082427040163, 082427040199, 082427040187, 082427040155, 082427040153, 082427040154, 082427040143, 082427040123, 082427040151, 082427040141, 082427040224, 082427040157, 082427220040, 082427040148, 082427140107, 082427010056, 082427080068, 082427100184, 082427160114, 082427250070, 082427140176, 082427310077, 082427160118, 082427100311, 082427110080, 082427090169, 082427060042, 082427240082, 082427210226, 082427150073, 082427150075, 082427150058, 082427150077, 082427160128, 082427290031, 082427160151, 082427210136, 082427070075, 082427070176, 082427070212, 082427070079, 082427010073, 082427090241, 082427090298, 082427090269, 082427210068, 082427190118, 082427120141, 082427120107, 082427120132, 082427120145, 082427120101, 082427120092, 082427200076, 082427200110, 082427020075, 082427020039, 082427030127, 082427030100, 082427030076, 082427210116, 082427230051, 082427140106, 082427140076, 082427130089, 082427130078, 082427030092, 082427310062, 082427130079, 082427130080, 082427130075, 082427130072, 082427130096, 082427290023, 082427230106, 082427040205, 082427040249, 082427040208, 082427040180, 082427040177, 082427040189, 082427040174, 082427040166, 082427040172, 082427040165, 082427040137, 082427040133, 082427040132, 082427040127, 082427040120, 082427040119, 082427040116, 082427120150, 082427180027, 082427200058, 082427150070, 082427150047, 082427110084, 082427020042, 082427010081, 082427050037, 082427050034, 082427050028, 082427110058, 082427090283, 082427160129, 082427160088, 082427090201, 082427120097, 082427140130, 082427180031, 082427180029, 082427180022, 082427210121, 082427030094, 082427150090, 082427100215, 082427170029, 082427160239, 082427140098, 082427100223, 082427100339, 082427100335, 082427150082, 082427140122, 082427210081, 082427310057, 082427120129, 082427090308, 082427090311, 082427090261, 082427090148, 082427020056, 082427190111, 082427100319, 082427090256, 082427060106, 082427260005, 0824270800
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenographe Pristina 3DNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDAOptima XR240amx, Mobile Digital Radiographic SystemJuly 22, 2026
- FDA3.0T SIGNA HDxt Family, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0,August 14, 2024
- FDADiscovery MR750w 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.5, DV29.1,August 14, 2024
Questions about this recall
Was Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System recalled?
Yes. FDA published recall Z-2464-2024 on August 14, 2024. Reported hazard: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
What should I do if I have this product?
Voluntary: Firm initiated · Letter