Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.
Published September 9, 2015 · Recall initiated: May 19, 2015
Summary
FDA records medical device Class II recall Z-2487-2015, published September 9, 2015 and initiated May 19, 2015. Recalling firm: Pointe Scientific, Inc.; Status: Terminated; Reported quantity: 2772 total vials (374 kits of 6 vials, 528 invidually sold vials); Lot: Lot: 505801 Expiration date: 2015-11 catalog numbers: G7583-CTL 7-G7583-CTL-L1 7-G7583-CTL-L2 7-G7583-CTL-L3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Pointe Scientific, Inc.
- Distributor
- Worldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.
- Lot
- Lot: 505801 Expiration date: 2015-11 catalog numbers: G7583-CTL 7-G7583-CTL-L1 7-G7583-CTL-L2 7-G7583-CTL-L3
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFIRMap 60mm Catheter; Size: 60 mm; Length 50.4 inches French Size Diameter: 8.5F (2.84mm) Model number :September 9, 2015
- FDAKimVent* Turbo-Cleaning Closed Suction System for Adults, 10 F, T-Piece (Product Code 227105) ASeptember 9, 2015
- FDAWestbrook's Hot Dilly Beans, Net WT. 12 oz (354.882G) glass jars, Ingredients: Green Beans, White DistilledSeptember 9, 2015
- FDANovaPlus Trach Care* Closed Suction System for Adults, Double Swivel Elbow, Directional Tip, 14F (Product CodeSeptember 9, 2015
- FDAWESTBROOK'S CANNERY HOT DILL PICKLES, Net WT. 16 oz (453.592g) glass jars, Ingredients: White DistilledSeptember 9, 2015
- FDAMonoject 0.9% Saline Flush Prefill Double Pouch Item Code: 8881570129 Product Usage: Sterile Delivery areSeptember 9, 2015
Questions about this recall
Was Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood. recalled?
Yes. FDA published recall Z-2487-2015 on September 9, 2015. Reported hazard: A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail