FDAFDA · Food and drugs·Recall number Z-2487-2015
Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.
Published September 9, 2015 · Recall initiated: May 19, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Pointe Scientific, Inc.
- Distributor
- Worldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.
- Lot
- Lot: 505801 Expiration date: 2015-11 catalog numbers: G7583-CTL 7-G7583-CTL-L1 7-G7583-CTL-L2 7-G7583-CTL-L3
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States